e-space
Manchester Metropolitan University's Research Repository

    The direct observation of magnetic images in electromagnet vibrating sample magnetometers

    Hoon, Steve R. and Willcock, S N M (1988) The direct observation of magnetic images in electromagnet vibrating sample magnetometers. Journal of physics e: scientific instruments, 21 (5). pp. 480-487. ISSN 0022-3735

    File not available for download.

    Abstract

    The authors report incontrovertible evidence of the first direct observation of ferromagnetic images in iron-core electromagnet vibrating sample magnetometers (VSMs) and determine their magnitude. They demonstrate experimentally that the ferromagnetic image strength at low field approaches that of the magnetic sample source moment and find image contributions of up to 25% in one typical VSM system studied. Quadrature image formation in nonmagnetic (Cu) plates in the proximity of the detection coils and pole pieces is also investigated. As a consequence of the observations reported it is clear that failure to account for images in electromagnet VSMs can lead to underestimating the induced signal by up to 100%.

    Impact and Reach

    Statistics

    Activity Overview
    6 month trend
    0Downloads
    6 month trend
    383Hits

    Additional statistics for this dataset are available via IRStats2.

    Altmetric

    Repository staff only

    Edit record Edit record